-40%
A&D AD-4212B-23 Production Weighing System, Lab Balance 21 g x 0.001 mg
$ 6863.47
- Description
- Size Guide
Description
To improve productivity, you can create your own weighing system using the AD-4212B Precision Weighing Sensor. The compact design, as well as the separate weighing unit and display unit, ensure installation of the Super Hybrid Sensor (SHS) at any stage of your weighing system.Weighing Capacity
21 g
Minimum Weighing Value
0.001 mg
Repeatability (Standard Deviation)
0 - 10 g (gross): 0.003 mg*
10 - 21 g (gross): 0.004 mg*
Linearity
±0.01 mg
Stabilization Time
Approximately 12 seconds (when set to FAST under a good environment)
Sensitivity Drift
±2 ppm / °C (10°C to 30° / 50°F to 86°F)
Operating Environment
5° C to 40° C (41° F to 104° F), 85% RH or less (no condensation)
Display Refresh Rate
10 times/second (5 times/second can be selected)
Display Mode
g, mg, oz, ozt, ct, mom, dwt, GN, tael, tola, PC (counting mode), Pct (percent mode)
Counting Mode
Minimum unit mass: 0.1 mg
Number of samples: 10, 25, 50 or 100 pieces
Percent Mode
Minimum 100% reference mass: 1 mg
Minimum 100% display: 0.01%, 0.1%, 1% (depends on the reference mass stored)
I/O Unit
RS-232C: bi-directional, 600-19200 bps
Comparator contact output: 3-level or 5-level (with buzzer)
External contact input: RE-ZERO, PRINT
Calibration Weight Provided(Conforming to OIML Class E2)
1 g
Applicable Calibration Weight Value
1 g, 2 g, 5 g, 10 g, 20g
Weighing Unit
Dimensions: 80 W x 230 D x 90 H mm
Weighing pan: ø 34 mm
Weight: approximately 2.3 kg
Display Unit
Dimensions: 237 W x 150 D x 155 H mm
Weight: approximately 2 kg
Connection Cable
Approximately 2 m (between weighing unit and display unit)
Power Supply
AC adapter
Power Consumption
Approximately 15 VA (supplied to the AC adapter)
* Repeatability for a 1 g mass using an automatic loading machine.